Technical parameters:
Accuracy: ±3% (infusion set of national standard after correct calibraion)
Applicable IV set: 15,20,60drops/ml, tube diameter: 3.4-4.5mm
Infusion rate: 0.1-1200ml/h, increment selectable: 0.1ml/h, 1ml/h, 10ml/h or 100ml/h
VTBI: 1-9999ml or 0 (ie. no limit on VTBI), increment selectable: 1ml/h, 10ml/h, 100ml/h or 1000ml/h
Volume infused: 0.0-36000ml
KVO rate: 1-5ml/, selectable by the user, default: 1ml/h
Bolus rate: 300-1200ml/h, selectable by the user, default: 1200ml/h
Purge rate: 600ml/h
Air bubble detection: Size of air bubble for filter: 50ul (1,2,3,4 levels adjustable, default: level 1)
Occlusion pressur: 40-160kpa, 3 levels adjustable(low, medium, high)
Waterproof level: IPX3
AC power: 100-240V 50/60Hz 25VA,
Built-in battery: Li_Polyme 7.4V 2100mAh; charge with power on: 10h, charge with power off: 3h, run 4h at 25ml/h after being fully charged.
Power consumption: 25VA
On-board battery/DC: DC 12V ±1V
Working conditions: 5C – 40C
Relative humidity: 10-90% (No condensation)
Storage temperature: -20C – 45C
Relative humidity: 10-95% (No condensation)
Shell size: 145(w)x 120(h)x 100(d)mm, not including pole clamp)
Net weight: <1.4kg